The TNT 12000 X-Ray Test Tools system is the newest and most comprehensive family of instruments available for assuring quality and safety of diagnostic x-ray imaging systems. With selection of all-in-one-exposure solid-state detector, dosimeter, ion chambers, optional mA/mAs invasive shunt or non-invasive clamp device, and choice of handheld display or laptop interface (both completely wireless), the TNT 12000 X-Ray Test Tools provide state-of-the-art solutions for any x-ray safety and analysis test protocol. A long battery life ensures uninterrupted operation all day.
The 12000WD solid-state detector sets up in seconds and measures kVp, radiation dose, dose rate, time, and half value layer (HVL) in a single exposure. The companion TNT 12000 DoseMate dosimeter and ion chambers provide precision dose and dose-rate measurement of radiographic, dental, fluoroscopic, and CT imaging systems. The wireless ZigBee® interface allows for quick testing and reporting, and at less than 1 mW power is preferred over Bluetooth® in medical settings where interference with delicate patient monitoring and treatment equipment may be a concern.
TNT 12000WD Wireless Detector Featuring all-in-one-exposure measurements and ZigBee wireless communication combined with the rugged, reliable, and accurate design that is a Fluke trademark, the TNT 12000WD is a new breed of non-invasive x-ray test tool. Its compact design enhances portability and wireless operation ensures setup in seconds.
Because the TNT 12000WD measures all parameters with every exposure, there is no need for complicated menu selection, further enhancing user productivity. TNT 12000WD always defaults to the last use when powered on, so when used often for repetitive procedures it is truly a one-button (power on) solution. TNT 12000WD has the expanded functionality needed for modern applications and can be managed with minimum keystrokes. Users can identify and select custom measurement protocols and save them for future use. Full test automation and documentation software is available for TNT 12000WD, creating the advantage of accurate, repeatable testing processes.
TNT 12000 DoseMate dosimeter The TNT 12000 DoseMate with ion-chambers provides the precision dose measurements needed for absolute dose measurement integrity. Offering customizable measurement protocols, the DoseMate dosimeter provides comprehensive, repeatable testing for radiographic, dental, fluoroscopic, and CT imaging systems in just a few keystrokes. The TNT 12000 DoseMate is compatible with existing Triad and NERO external ion chambers, offering the newest dosimeter technology with minimum investment.
mA/mAs measurement Integrated mA/mAs measurement capability is available with both the TNT 12000WD and the TNT 12000 DoseMate. A mA/mAs-standard invasive shunt or optional non-invasive clamp device are available to meet the challenges of mAs measurement with all types of x-ray systems. ZigBee wireless interface delivers measured results to the user’s choice of handheld display or laptop, integrating all QA measurements on a single readout.
TNT 12000D Wireless Display The TNT 12000D Wireless Display has been specifically designed with user productivity in mind and displays all measured results, as well as identification information, battery level, calibration date and more. ZigBee wireless interface from the test device offers targeted communication to the mating display for speedy results reading up to 100 feet from the test device. Wireless communication and intelligent device interface allows the user to perform multiple exposures without leaving the radiation-shielded area, thus allowing complete test protocols to be performed with minimum travel between operator and test device. Low power requirements for ZigBee interface allow for extensive battery life as compared to Bluetooth test devices, which require frequent recharging from inefficient power drain.
Ansur Test Automation Software Test automation software allows preventative maintenance routines to be tightly controlled and consistently reported while directing users through prescribed test protocols with simple and efficient instruction. Test standardization minimizes human error and alleviates training requirements with step-by-step instruction. Photos, drawings and instructional visual aids included in test protocols negate the need to carry bulky service manuals for complicated PM procedures. Professional electronic reporting and database/CMMS interface allows for total PM and service data management.
Choice of all-in-one detector, dosimeter, integral ma/mAs and handheld display or user’s own laptop interface.
100 % ZigBee wireless operation
Compact handheld design
Sets up in seconds
Displays all values in one shot (all-in-one detector)
Simple user interface
Unbeatable ruggedness for long-term reliability
40 kHz kV sampling rate to ensure accuracy with the most difficult applications (TNT 12000WD)
Global support network delivering prompt service and peace of mind to Fluke Biomedical customers worldwide
TNT 12000 X-Ray Test Tools
Physical specifications Display 320 X 240 Color LCD Size (WxDxH) TNT 12000D Display 6.1 in x 4.5 I x 1.6 in TNT 12000WD Wireless Detector 6.7 in x 4.5 in x 1.75 in TNT 12000 DoseMate Dosimeter 6.7 in x 4.5 in x 1.75 in Weight TNT 12000D Display 0.422 kg (0.93 lb) TNT 12000WD Wireless Detector 1.5 lb TNT 12000 DoseMate Dosimeter 1.1 lb
Electrical specifications Battery Type Lithium-Ion, 3.7 V, 4000 mAh Charge time Approximately 5 hr Operating duration Approximately 8 hr minimum AC adapter Input voltage 100 V ac to 240 V ac Input frequency 50/60 Hz Input current 0.5 A (rms) Output voltage 6 V dc
Environmental specifications Operating temperature 0 ºC to 35 ºC (32 ºF to 122 ºF) Storage temperature -35 ºC to 50 ºC (-31 ºF to 122 ºF) Operating humidity 20 % to 80 % RH (non-condensing)
Measurements kVp Units kVp Average (average of peaks during a specified interval) kVp Max (highest peak during a specified interval) PPV (peak practical voltage) Ranges Radio/Fluoro modes: 40 kV to 150 kV Mammo modes: Mo/Mo: 22 kV to 35 kV (standard calibration) Rh/Rh: 25 kV to 49 kV (optional calibration) Mo/Rh: 22 kV to 40 kV (optional calibration) Mo/Al: 22 kV to 49 kV (optional calibration) Rh/Al: 22 kV to 49 kV (optional calibration) W/Rh: (optional calibration) W/Ag: (optional calibration) Resolution 0.1 kV Accuracy Radio/Fluoro modes: ± 2 % or ± 1 kV, whichever is greater Mammo modes: ± 2 % or ± 0.7 kV, whichever is greater Reproducibility ± 1 % (std of 5 readings) Filtration correction range Radio/Fluoro modes: 1 mm Al to 10 mm Al or equivalent Mammo mode: 0 mm Al to 0.4 mm Al added filtration Dose/Exposure Units Roentgens, Grays Range 0.5 mR to 999 R 5 µGy to 999 Gy Resolution 1 µR 0.01 µGy Accuracy ± 5 % Reproducibility ± 0.5 % (std of 5 readings) Filtration correction range Radio/Fluoro modes: 1 mm Al to 10 mm Al or equivalent Mammo modes: 0 mm Al to 0.4 mm Al added filtration kV correction range Radio/Fluoro modes: 40 kV to 150 kV Mammo modes: Mo/Mo: 22 kV to 35 kV Dose/Exposure rate Units Roentgens or Grays per hour, minute, second, pulse Range 8 mR/s to 10 R/s 70 µGy to 100 mGy/s 130 µR/pulse to 160 mR/pulse (@ 60 PPS) 1.2 µGy/pulse to 1.4 mGy/pulse (@ 60 PPS) Resolution 1 µR/s 0.01 µGy/s 0.02 µR/pulse (@ 60 PPS) 0.2 n Gy/pulse (@ 60 PPS) Accuracy ± 5 % Filtration Correction Range Radio/Fluoro Modes: 1 mm Al to 10 mm Al or equivalent Mammo Mode: 0 mm Al to 0.4 mm Al added filtration kV Correction Range Radio/Fluoro Modes: 40 kV to 150 kV Mammo Mode: Mo/Mo: 22 kV to 35 kV Exposure time–radiographic modes Units Milliseconds, pulses Range Range (@ stated accuracy) Milliseconds: 10 ms to 9999 ms Pulses: 1 pulse to 999 pulses Resolution Milliseconds: 0.1 ms Pulses: 1 pulse Accuracy Milliseconds: 1 % or 0.5 ms Pulses: ± 1 pulse Reproducibility Milliseconds: 1 % or 0.5 ms Pulses: ± 1 pulse Elapsed time–fluoro modes Range 10 sec to 9999 sec Resolution 0.1 second Accuracy 1 % or 0.5 sec Average pulse rate – pulsed fluoro Range 1 pps to 999 pps (pulses per second) Resolution 1 pps Accuracy 1 pps Average pulse width – pulsed fluoro Range 10 ms to 999 ms Resolution 0.1 ms Accuracy 1 % or 0.5 ms HVL Range Radio/Fluoro modes: 1.2 mm Al to 10 mm Al (equivalent) Mammo mode: 0.2 mm Al to 0.6 mm Al (equivalent) Resolution Radio/Fluoro modes: 0.1 mm Al (equivalent) Mammo mode: 0.01 mm Al (equivalent) Accuracy Radio/Fluoro modes: ± 10 % or 0.2 mm Al (equivalent) Mammo mode: ± 5 % or 0.05 mm Al (equivalent)
TNT 12000 DoseMate dosimeter specifications
Accuracy Exposure and exposure rate accuracy ± 1 % of reading ± 2 range resolution steps over range of 18 °C to 28 °C and ± 2 % of reading ± 2 range resolution steps over the full operating temperature range of 0 °C to 50 °C 3 % NIST-traceable calibration is provided with each system and includes effects of 96035B, and 96020C
Measurement modes Dose Autoranging across five decades of sensitive ranges Automatic drift and offset compensation Automatic post-exposure display hold Rate Measurement range covers a span from low-level image intensifier measurements to unattenuated, direct beams Automatic offset compensation and nonlinear filtering Autoranging provides five decades of sensitivity ranges Display updates once per second
Low Rate Mode This mode is only for making very low dose rate measurements. Nonlinear digital filtering and autoranging provide five decades of sensitivity ranges. Display updates once per second. In this mode, automatic current offset and drift compensation are disabled. As a result, the system can display very low dose rates
Power requirements Bias voltage supply Fixed electronic bias (~300 V) Bias voltage removed from triaxial input connector at instrument turnoff
Connections Ion chamber input Triax, BNC Collector and guard positive-biased relative to ion chamber body and service dosimeter chassis Power Same as TNT 12000WD
General information Display Wireless ZigBee or optional USB interface to TNT 12000 display indicates all ion chamber identification information, numerical measurement results, battery level, calibration date, and other information Weight ? 0.68 kg (1.5 lb) Dimensions (maximum) (WxDxH) 15.2 cm x 11.4 cm x 4.45 cm (6 in x 4.5 in x 1.75 in)
96020C and 96035B Diagnostic Ion Chambers
Technical specifications Energy range 96020C 30 kVp to 150 kVp 96035B 30 kVp to 150 kVp for diagnostic measurements 20 kVp to 50 kVp for mammographic measurements Nominal volume 96020C 150 cm3; 11.30 cm diameter by 1.5 cm thick active volume 96035B 15 cm3; 3.96 cm diameter by 1.22 cm thick active volume Nominal sensitivity 96020C H60: 2.08 R/C x 107 R/C at 22 °C and 760 mmHg (optimized for low-level image intensifier and cine measurements) 96035B L100: 2.0 R/C x 108 R/C at 22 °C and 760 mmHg MV30 (PTB mammo point) 96020C/96035B 2.21 R/C x 108 R/C at 22 °C and 760 mmHg (flat energy response suitable for conventional diagnostic radiography and mammography) Leakage current 96020C/96035B < 10 fA under normal bias conditions (300 V) Collection efficiency 96020C 95 % at 2,000 R/min 96035B 95 % at 5,000 R/min Wall material 96020C Composite graphite-filled thermoplastic 96035B Graphite-coated acrylic (methylmethacrylate) Window material 96020C 0.76 mm thick, graphite-coated polycarbonate 96035B Both entrance windows are made of 0.25 mm graphite-coated polycarbonate Window density 96020C 91 mg/cm2 96035B 32 mg/cm2 Active window area 96020C 100 cm2, centered within the chamber body 96035B Each side of the chamber has a circular active window region centered 7.1 mm further from the BNC connector than the center of the chamber body; active window regions have an area of 12.32 cm2 Collector plate 96020C 0.8 mm thick graphite-coated acrylic plate, 10.80 cm in diameter; 2.16 cm x 2.85 cm guard region electrically isolated from collector area 96035B 0.25 mm thick, centrally mounted, graphite-coated, polycarbonate plate, 3.18 cm, ± 0.01 cm in diameter; 1.27 cm x 0.89 cm guard region is electrically isolated from the collector area Connector 96020C/96035B Side-mounted, triaxial, two-lug BNC connector Calibration 96020C Standard calibration performed at H60 (NIST defined as 60 kVp, first HVL of 6 mm Al, homogeneity coefficient of 94) 96035B Standard calibration performed at one diagnostic and one mammographic beam quality Calibration factors normalized to 22 °C and 760 mmHg Diagnostic Unattenuated Beam Calibration on diagnostic side of chamber is performed at M80 (NIST defined as 80 kVp, first HVL of 2.97 mm Al, homogeneity coefficient of 57) Mammographic Beam Calibration on mammographic side performed at Mo/Mo28 (NIST defined as 28 kVp, first HVL of 0.332 mm Al, homogeneity coefficient of 74.3) or MV30 (PTB defined as 30 kVp, first HVL of 0.337 mm Al)
mA/mAs specifications
Measurement range Invasive mAs (with shunt) mA/mAs ranges* Min to Max 0.00 to 99.99 100.0 to 999.9 1000 to 1999 *mAs measurements are fully auto-ranging, so values are display ranges only Non-invasive mAs (with clamp) mA/mAs ranges Min to Max 0.0 to 999.9 1000 to 3999
Accuracy Invasive mAs (with shunt) ± 1 % Model Name Description Price* TNT 12000WD Wireless Detector + Display TNT 12000WD Wireless Detector + TNT 12000D Display (Item # 3335763)